Pressure Controllers
PPC4
|
| Ranges |
Accuracy (Uncertainty) |
2
200/700/2000 kPa
3
3500/7000 kPa
10
10000 kPa
(vacuum available as well) |
±0.1% FS
or - using Q-RPT modules -
±0.0008%FS
±0.0048% FS
(0
30% of range)
±0.009% RDG
±0.016% FS
(30
100% of range)
|
|
|
PPC4E
|
| Ranges |
Accuracy (Uncertainty) |
-Ðatm…14 MPa gauge, 0…14 MPa abs.
-Ðatm…7 MPa gauge, 0…7 MPa abs.
-Ðatm…1,4 MPa gauge, 0…1,4 MPa abs.
-Ðatm…100 kPa gauge, 0…200 kPa abs.
±15 kPa gauge
|
from ±0,02%FS of sub-range |
|
PPCH-G
|
| Ranges |
Accuracy (Uncertainty) |
0,7 … 20/40/70/100 MPa
0,7 … 7/10/14/20/40/70/100 MPa
|
±0.1% FS
or - using Q-RPT modules -
±0.004 … ±0.005% FS
(0 … 30% of range) and
±0.013 … ±0.016% RDG
(30 … 100% of range)
|
|
PPCH
|
| Ranges |
Accuracy (Uncertainty) |
1
70/100/140/200 MPa
1
20/40/70/100/140/200 MPa |
±0.1% FS
or - using Standard class Q-RPT -
±0.0048
±0.006% FS
(0
30% of range) and
±0.016
±0.020% RDG
(30
100% of range)
|
|
Built-in pneumatically driven hydraulic pump
|
|
RPM4 Reference Pressure Monitor
|
| Ranges |
Accuracy (Uncertainty) |
10/15/60/100/200/250 kPa gauge 110/160/200/350 kPa abs. 0.7/1.4/2/3.5/7/10 MPa abs., gauge |
±0.008
±0.01% RDG |
| 14/20/40/70/100/140 MPa abs., gauge |
±0.013% RDG |
| 200/280 MPa abs., gauge |
±0.018% RDG |
|
|
|
E-DWT-H electronic deadweight tester
|
| Ranges |
Accuracy (Uncertainty) |
|
7/10/14/20/40/70/100/140/200 gauge
|
±0.02% RDG
|
|
|