AboutNewsProductsDocumentationInquiries
Search:
AnalyzersCalibratorsTest BenchesFlow Meters
   Multifunction   Pressure   Temperature   Electric signals   Software

MC5-R (MC5P-R)

Brochure | Order List
Calibrator MC5-R
Measurement/generation of electric signals and frequency
Ranges Accuracy
±50 V/12 V ±(0.02% RDG + 0.25 mV/0.1 mV)
±1000 mV/±500 mV ±(0.02% RDG + 5 µV/4 µV)
±100 mA/±25 mA ±(0.02% RDG + 1.5 µA/1 µA)
0/1...4000 Ohm ±(0.02% RDG + 3.5 mOhm)/±0.04% RDG
0.0028...50000 Hz ±0.01% RDG
0...9 999 999 pulses
Extra temperature curves, complying with the following GOST standards
GOST 3044-84, GOST R 50431-92, GOST R 8.585-2001,
GOST 6651-78, GOST 6651-84, GOST 6651-94, GOST R 8.625-2006

Pressure measurement (internal and external modules)
Ranges Accuracy
6 months 12 months
±1 kPa diff. ±(0.06% RDG + 0.035% Span) ±(0.1% RDG + 0.05% Span)
0...10 kPa ±(0.015% RDG + 0.017% FS) ±(0.025% RDG + 0.025% FS)
±40 kPa ±(0.015% RDG + 0.015% FS) ±(0.025% RDG + 0.02% FS)
±100 kPa ±(0.015% RDG + 0.01% FS) ±(0.025% RDG + 0.01% FS)
-0.1...+0.2/0.6/2/6/10/16 MPa ±(0.015% RDG + 0.007% FS) ±(0.025% RDG + 0.01% FS)
0...16/25/60/100 MPa ±(0.015% RDG + 0.01% FS) ±(0.025% RDG + 0.015% FS)

Optional documenting capability using QD3 or CMX software

Optional HART and/or Foundation Fieldbus H1 / Profibus PA communication available as well as special MC5P-R panel mounted version


MC5-R-IS

Brochure | Order List
Calibrator MC5-R-IS
Measurement/generation of electric signals and frequency
Ranges Accuracy
±30 V/-2.5...12 V ±(0.02% RDG + 0.2 mV/0.1 mV)
±250 mV ±(0.02% RDG + 4 µV)
±100 mA/0...25 mA ±(0.02% RDG + 1.5 µA/1 µA)
0/1...4000 Ohm ±(0.02% RDG + 3.5 mOhm)/±0.04% RDG
0.0028...50000 Hz ±0.01% RDG
0...9 999 999 pulses
Extra temperature curves, complying with GOST standards
GOST 3044-84, GOST R 50431-92, GOST R 8.585-2001,
GOST 6651-78, GOST 6651-84, GOST 6651-94, GOST R 8.625-2006

Pressure measurement (internal and external modules)
Ranges Accuracy
6 months 12 months
±1 kPa diff. ±(0,06% RDG + 0,035% Span) ±(0,1% RDG + 0,05% Span)
0...10 kPa ±(0,015% RDG + 0,017% FS) ±(0,025% RDG + 0,025% FS)
±40 kPa ±(0,015% RDG + 0,015% FS) ±(0,025% RDG + 0,02% FS)
±100 kPa ±(0,015% RDG + 0,01% FS) ±(0,025% RDG + 0,01% FS)
-0.1...+0.2/0.6/2/6/10/16 MPa ±(0,015% RDG + 0,007% FS) ±(0,025% RDG + 0,01% FS)
0...16/25/60/100 MPa ±(0,015% RDG + 0,01% FS) ±(0,025% RDG + 0,015% FS)

Explosion proof marking 0ExiaIICT4 X

Optional documenting capability using QD3 or CMX software


MC4-R

Brochure | Order List
Calibrator MC4-R
Measurement of electric signals and frequency
Ranges Accuracy
±0.25/±1000 V ±(0.02% RDG + 5 µV)
1 … 25/60 V ±(0.02% RDG + 0.25 mV)
±25/±100 mA ±(0.02% RDG + 1.5 µA)
0 … 250/2650/4000 Ohm ±(0.02% RDG + 3.5 mOhm)
0.0028 … 50000 Hz ±0.01% RDG
0 … 9 999 999 pulses  

Generation of electric signals and frequency
Ranges Accuracy
-25 … 150 mV ±(0.02% RDG + 4 µV)
±0.25 V ±(0.02% RDG + 0.1 mV)
-3 … 12 V ±(0.02% RDG + 0.1 mV)
0 … 25 mA ±(0.02% RDG + 1.5 µA)
0 … 400/4000 Ohm ±0.04% RDG or 30 mOhm
0.0005 … 10000 Hz ±0.01% RDG
0...9 999 999 pulses  
Extra temperature curves, complying with the following GOST standards:
GOST 3044-84, GOST R 50431-92, GOST R 8.585-2001,
GOST 6651-78, GOST 6651-84, GOST 6651-94, GOST R 8.625-2006

Pressure measurement
Ranges Accuracy
6 months 12 months
Internal and external modules:
±20 kPa
-0.1 … 0.2/2 MPa, 0 … 16 MPa

±(0.03% RDG + 0.017% FS)
±(0.025% RDG + 0.01% FS)

±(0.05% RDG + 0.025% FS)
±(0.035% RDG + 0.015% FS)
External modules:
±1 kPa diff.
0 … 10 kPa
±40 kPa
±100 kPa
-0.1 … +0.2/0.6/2/6/10/16 MPa
0 … 25/60/100 MPa

±(0.06% RDG + 0.035% FS)
±(0.015% RDG + 0.017% FS)
±(0.015% RDG + 0.015% FS)
±(0.015% RDG + 0.01% FS)
±(0.015% RDG + 0.007% FS)
±(0.015% RDG + 0.01% FS)

±(0.1% RDG + 0.05% FS)
±(0.025% RDG + 0.025% FS)
±(0.025% RDG + 0.02% FS)
±(0.025% RDG + 0.01% FS)
±(0.025% RDG + 0.01% FS)
±(0.025% RDG + 0.015% FS)

Optional documenting capability using CMX software

MC2-R

Brochure | Order List
Calibrator MC2-R
Measurement of electric signals and frequency
Ranges Accuracy
±0.25 … ±1 V ±(0.02% RDG + 5 µV)
1 … 25/60 V ±(0.02% RDG + 0.25 mV)
±25/±100 mA ±(0.02% RDG + 1.5 µA)
0 … 250/2650/4000 Ohm ±(0.02% RDG + 3.5 mOhm)
0.0028 … 50000 Hz ±0.01% RDG
0 … 9 999 999 pulses  

Generation of electric signals and frequency
Ranges Accuracy
-25 … 150 mV ±(0.02% RDG + 4 µV)
±0.25 V ±(0.02% RDG + 0.1 mV)
-3 … 12 V ±(0.02% RDG + 0.1 mV)
0 … 25 mA ±(0.02% RDG + 1.5 µA)
0 … 400/4000 Ohm ±0.04% RDG or 30 mOhm
0.0005 … 10000 Hz ±0.01% RDG
0...9 999 999 pulses  
Extra temperature curves. complying with the following GOST standards:
GOST 3044-84, GOST R 50431-92, GOST R 8.585-2001,
GOST 6651-78, GOST 6651-84, GOST 6651-94, GOST R 8.625-2006

Pressure measurement
Ranges Accuracy
6 months 12 months
Internal and external modules:
±20 kPa
-0.1 … 0.2/2 MPa, 0 … 16 MPa

±(0.03% RDG + 0.03% FS)
±0.035% FS

±(0.05% RDG + 0.05% FS)
±0.05% FS
External modules:
±1 kPa diff.
0 … 10 kPa
±40 kPa
±100 kPa
-0.1 … +0.2/0.6/2/6/10/16 MPa
0 … 16/25/60/100 MPa

±(0.06% RDG + 0.035% FS)
±(0.015% RDG + 0.017% FS)
±(0.015% RDG + 0.015% FS)
±(0.015% RDG + 0.01% FS)
±(0.015% RDG + 0.007% FS)
±(0.015% RDG + 0.01% FS)

±(0.1% RDG + 0.05% FS)
±(0.04% RDG + 0.025% FS)
±(0.04% RDG + 0.02% FS)
±(0.025% RDG + 0.01% FS)
±(0.04% RDG + 0.01% FS)
±(0.04% RDG + 0.015% FS)

MC2-R-IS

Brochure | Order List
Calibrator MC2-R-IS
Measurement of electric signals and frequency
Ranges Accuracy
±0.25 … ±1 V ±(0.02% RDG + 5 µV)
1 … 25/60 V ±(0.02% RDG + 0.25 mV)
±25/±100 mA ±(0.02% RDG + 1.5 µA)
0 … 250/2650/4000 Ohm ±(0.02% RDG + 3.5 mOhm)
0.0028 … 50000 Hz ±0.01% RDG
0 … 9 999 999 pulses  

Generation of electric signals and frequency
Ranges Accuracy
-25 … 150 mV ±(0.02% RDG + 4 µV)
±0.25 V ±(0.02% RDG + 0.1 mV)
-3 … 12 V ±(0.02% RDG + 0.1 mV)
0 … 25 mA ±(0.02% RDG + 1.5 µA)
0 … 400/4000 Ohm ±0.04% RDG or 30 mOhm
0.0005 … 10000 Hz ±0.01% RDG
0...9 999 999 pulses  
Extra temperature curves. complying with the following GOST standards:
GOST 3044-84, GOST R 50431-92, GOST R 8.585-2001,
GOST 6651-78, GOST 6651-84, GOST 6651-94, GOST R 8.625-2006

Pressure measurement (external modules)
Ranges Accuracy
6 months 12 months
EXT-s-IS modules
±20 kPa
-0.1 … 0.2/2 MPa, 0 … 16 MPa

±(0.03% RDG + 0.03% FS)
±0.035% FS

±(0.05% RDG + 0.05% FS)
±0.05% FS
EXT-IS modules
±1 kPa diff.
0 … 10 kPa
±40 kPa
±100 kPa
-0.1 … +0.2/0.6/2/6/10/16 MPa
0 … 25/60/100 MPa

±(0.06% RDG + 0.035% Span)
±(0.015% RDG + 0.017% FS)
±(0.015% RDG + 0.015% FS)
±(0.015% RDG + 0.01% FS)
±(0.015% RDG + 0.007% FS)
±(0.015% RDG + 0.01% FS)

±(0.1% RDG + 0.05% Span)
±(0.025% RDG + 0.025% FS)
±(0.025% RDG + 0.02% FS)
±(0.025% RDG + 0.01% FS)
±(0.025% RDG + 0.01% FS)
±(0.025% RDG + 0.015% FS)

Explosion proof marking 0ExiaIICT4 X

ASC300-R

Brochure | Order List

Measurement/generation of electric signals
Ranges Accuracy
0 … 30/20 V ±(0.015% RDG + 2 mV)
-10 … 75 mV ±(0.02% RDG + 10 µV)
0 … 24 mA ±(0.015% RDG + 2 µA)
5 … 4000 Ohm ±(0.025% RDG + 3.5 mOhm)
1 … 10000 Hz ±(0.05% RDG + 0.1 Hz)
2 … 600 pulse/min
(1 … 30000 pulse sequence generation)
±(0.05% RDG + 0.1 pulse/min)
Extra temperature curves, complying with the following GOST standards:
GOST 3044-84, GOST R 50431-92, GOST R 8.585-2001, GOST 6651-78, GOST 6651-84, GOST 6651-94

Pressure measurement (external modules)
Ranges Accuracy
6 months 12 months
APM H modules:
0 … 7 MPa (gauge);
0 … 20 MPa (abs)
±0.01% FS ±0.025% FS
APM S modules:
±0.1; 0 … 70 MPa (gauge)
0 … 700 kPa (abs, diff)
±0.025% FS ±0.05% FS


 Copyright © 2001—2008 "Artvik"